Gas Chromatography

Control system: designed for monitoring and controlling the instrument via the computer.

Column Compartment/ oven with superior thermal performance, multistage (10 ramps) programmed temperature control function. (supported by“control system”)

Advanced built- in data acquisition system, supporting real time instrument status monitoring, detection signal acquisition and PC control

Column oven accommodates up to 3 chromatographic columns, and supports quick heat- up and rapid cool- down with automated back- door opening.

Flexible sample introduction system: 3 sample injectors could be installed and operated simultaneously with independent temperature control.

High sensibility and stability detector.

2 independent and analog signals output.

M6 software, compatible with GLP/FDA-21 CFR Part11 requirements and regulations.(electronic records and signatures)

Sample Introduction System sample injector and evaporation chamber
Column Oven

Flame Ionization Detector (FID)

Temperature range Ambient temperature +7℃~400℃  (in 1 ℃ increment)
Temperature accuracy ±0.02℃
Cooling time 400℃ to 50℃ in 8-10 min at 25℃ ambient
Programmed temperature setting 0.1℃~40℃/ min(in 1℃ increment)
Program ramps 7 ramps in total (10 ramps available with control workstation)
Size (L x W x H) 284 x 280 x 241 mm (inside) 340 x 345 x 281mm  (outside)
Flame Ionization Detector (FID) Detection limit 3×10-¹²g/s (C₁₆)
Best test result 3×10-¹²g/s (C₁₆)
Drift 6×10-13A/30 min
Noise ≤5×10–14A
Linearity range ≥ 10⁶
Thermal Conductivity Detector (TCD) Sensitivity ≥5000mV, mL/ mg(C16)
Baseline noise ≤20μV
Drift ≤60μV/h
Linearity range ≥104
Flame Photometric Detector (FPD) Detection limit ≤8×10-¹³g/s(P);  ≤8×10-¹¹g/s(S)
Drift 2×10-¹¹A/30min
Baseline noise 5×10-¹²A
Nitrogen-Phosphorus Detector (NPD) Detection limit ≤5×10-12g/s (N)  (Azobenzene)

≤5×10-13g/s (P)  (Malathion)

Electron Capture Detector (ECD) Linear range ≥103(γ-666)
Baseline drift ≤80μV/h
Baseline noise ≤20μV
Detection limit ≤2×10-13/ml(γ-666)